Fluke Corporation MET/CAL Procedure ============================================================================= INSTRUMENT: Fluke 83: CAL ADJ /5100 INSTRUMENT: Fluke 85: CAL ADJ /5100 INSTRUMENT: Fluke 87: CAL ADJ /5100 DATE: 07-Feb-95 AUTHOR: Fluke Corporation REVISION: $Revision: 2.2 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 1 NUMBER OF LINES: 74 CONFIGURATION: Fluke 5100B ============================================================================= # # Source: # Fluke 83,85,87 Service Manual. # (PN 834168 April 1989, Rev. 3, 7/93, Change/Errata 8/94). # # Compatibility: # MET/CAL 4.0 or later # # Subprocedures: # None # # Required Files: # 80_pca.bmp # # System Specifications: # TUR calculation is based on specification interval of the accuracy file. # The default 5100B accuracy file contains 90 day specs. # # Fluke makes no warranty, expressed or implied, as to the fitness # or suitability of this procedure in the customer's application. # # The 90 day specifications of the 5100B are used in TUR computations. # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- J F W 1.002 HEAD EQUIPMENT SETUP 1.003 DISP Disconnect the test leads if attached. Remove the three 1.003 DISP phillips-head screws from the bottom of the case. Turn 1.003 DISP the instrument face-up, and grasp the top cover at both 1.003 DISP sides of the input connectors. Then, pull the top cover 1.003 DISP from the UUT. 1.004 DISP Turn the UUT on. 1.004 DISP Ambient temperature: 18C - 28C. 1.004 DISP Relative humidity: less than 80%. 1.004 DISP Allow the UUT to stabilize. 1.005 HEAD DC VOLTAGE CALIBRATION 1.006 DISP Select the V DC function on the UUT. 1.007 DISP Connect the 5100B and the UUT as follows. 1.007 DISP [32] 5100B OUTPUT HI to UUT V Ohm Diode 1.007 DISP [32] 5100B OUTPUT LO to UUT COM 1.008 5100 3.5V S 2W 1.009 PIC 80_PCA (20,3,R/W) "Adjust R21 for a UUT reading of" 1.009 PIC (22,3,R/W) "3.500+/-0.001V" 1.010 5100 * S 1.011 HEAD AC VOLTAGE CALIBRATION 1.012 DISP Select the V AC function on the UUT. 1.013 JMPT 1.017 Fluke 87: CAL ADJ /5100 1.014 5100 3.5V 100H S 2W 1.015 PIC 80_PCA (2,3,R/W) "Adjust R34 for a UUT reading of" 1.015 PIC (22,3,R/W) "3.500+/-0.002V" 1.016 JMP 1.019 1.017 5100 3.513V 50H S 2W 1.018 PIC 80_PCA (20,3,R/W) "Adjust R34 for a UUT reading of" 1.018 PIC (22,3,R/W) "3.500+/-0.002V" 1.019 5100 3.5V 10kH S 2W 1.020 PIC 80_PCA (20,3,R/W) "Adjust C2 for a UUT reading of" 1.020 PIC (22,3,R/W) "3.500+/-0.004V" 1.021 5100 35V 10kH S 2W 1.022 PIC 80_PCA (20,3,R/W) "Adjust C3 for a UUT reading of" 1.022 PIC (22,3,R/W) "35.00+/-0.04V" 1.023 END